background image

Aeroneb

®

 Lab Nebulizer System Instruction Manual  

11 

 

 

Maintenance 

This section describes how to clean, sterilize, and inspect Aeroneb

® 

Lab components. 

 
Nebulizer unit (including filler cap).  

To clean unit between uses for the same subject 

1.  Disconnect the control module and cable from the nebulizer unit. 

2.  Remove the filler cap from the nebulizer unit. 

3.  Clean parts with warm water and mild liquid detergent.  Rinse thoroughly and air dry. 

CAUTION:

 Do not use abrasive or sharp tools to clean the nebulizer unit. 

4.  Check for cracks or damage, and replace if any defects are visible. 

5.  Place the disassembled components into appropriate sterilization wrapping. 

CAUTION:

 Do not reassemble parts prior to autoclaving. 

6.  To sterilize, autoclave wrapped parts using steam sterilization pre-vacuum cycle, 132

°

C -135

°

(270°F - 275°F) for 3 - 4 minutes with drying cycle. 

7.  Prior to next use, check for cracks or damage, and replace if any defects are visible. 

8.  Prior to next use, perform a functional test as described in this manual. 

NOTE:

   It is recommended that the device be cleaned after each use to ensure proper nebulization 

Control module, AC/DC adapter and cable. 

1.  Wipe clean with a damp cloth. 

2.  Do not autoclave.  

3.  Check for exposed wiring, damaged connectors, or other defects and replace if any are visible. 

4.  Perform functional test as described in this manual before re-use. 

 

Содержание Aeroneb lab AG-AL1000

Страница 1: ... 2004 Aerogen Inc Part No AG AL1010 Rev C Instruction Manual ...

Страница 2: ...cifications 13 Physical 13 Environmental 13 Performance 13 Power 13 List of Figures Figure 1 Aeroneb Lab 4 Figure 2 Aeroneb Lab controls and indicators 7 Figure 3 Connecting filler cap to nebulizer unit 8 Figure 4 Connecting nebulizer unit to control module 8 Figure 5 Filling the nebulizer unit 9 Figure 6 Maximum fill indication point 9 List of Tables Table 1 Aeroneb Lab System symbols 6 Table 2 A...

Страница 3: ...viding 9 volts DC This can enable the timing of aerosolization based on programmed criteria The Aeroneb Lab is a latex free nebulizer system that incorporates Aerogen s proprietary OnQ Aerosol Generator for aerosolization Note The Aeroneb Lab utilizing OnQ Aerosol Generator technology has been utilized to aerosolize a broad range of liquid formulations including suspensions and proteins However th...

Страница 4: ...t is an OnQ Aerosol Generator which consists of a domed aperture plate with precision formed holes that control the size of the aerosol droplets and a vibrational element whose micro pumping action aerosolizes liquid Gravity brings the liquid in contact with the aerosol generator where the liquid is drawn through the aperture plate and converted into an aerosol The control module can be powered by...

Страница 5: ...ice user Cautions To avoid mechanical or electrical damage do not drop the nebulizer unit or the control module Disconnect nebulizer unit from control module before cleaning Do not immerse in water or autoclave the control module cables or AC DC adapter Use only with components specified by Aerogen Inspect all parts before use and do not use if any parts are missing cracked or damaged In case of m...

Страница 6: ... Serial number where YY is the year of manufacture and XXXX is the serial number Control Module Output A C voltage Attention consult accompanying documents Control Module Input D C voltage Degree of protection against dripping water Fragile handle with care Class II equipment per IEC 61010 1 Storage temperature limitations 20 C to 60 C On off power Keep dry Output Compliance with the European Dire...

Страница 7: ...Figure 2 Aeroneb Lab controls and indicators Table 2 Aeroneb Lab control indicator functions Control indicator Function Amber Warning LED When illuminated indicates nebulizer is disconnected or faulty connection Green Power delivery LED When illuminated indicates that power is supplied and that the unit is aerosolizing ...

Страница 8: ...onnecting nebulizer unit to control module NOTE To ensure uninterrupted operation of the Aeroneb Lab secure power and control module cables so they cannot become disconnected during treatment Ensure all cables are routed safely NOTE To ensure proper nebulization maintain the nebulizer in an upright position 4 When powering the control module using the DC cable Insert one end of the DC cable into t...

Страница 9: ...the nebulizer Figure 5 3 Close the filler cap tab CAUTION To avoid damage to the nebulizer unit do not use a syringe with needle The maximum capacity of the nebulizer unit is 10 mL Do not fill the nebulizer unit beyond the maximum fill indication point Figure 6 The underside of the filler cap represents maximum fill indication point Figure 5 Filling the nebulizer unit MAXIMUM INDICATION FILL POINT...

Страница 10: ...he device for cracks or damage and replace if any defects are visible 2 Pour 0 5 mL of sterile water or normal saline into the nebulizer unit 3 Connect the nebulizer unit to the control module connect the control module to the power source 4 Verify that the green power light is on Ensure that aerosol is visible 5 Disconnect power to turn the system off 6 Discard any remaining liquid before next us...

Страница 11: ...lace the disassembled components into appropriate sterilization wrapping CAUTION Do not reassemble parts prior to autoclaving 6 To sterilize autoclave wrapped parts using steam sterilization pre vacuum cycle 132 C 135 C 270 F 275 F for 3 4 minutes with drying cycle 7 Prior to next use check for cracks or damage and replace if any defects are visible 8 Prior to next use perform a functional test as...

Страница 12: ... Aeroneb Lab parts list If these suggestions do not correct the problem discontinue use of any device that is or appears to be damaged or not operating properly and contact your local Aeroneb Lab product representative Order Numbers Table 4 lists the Aeroneb Lab order numbers see Figure 1 for pictures Table 4 Aeroneb Lab parts list Description Order Number Aeroneb Lab Control Module includes DC ca...

Страница 13: ...fied performance at temperatures up to 45 C 113 F Atmospheric pressure 450 to 1100 hPa Humidity 15 to 95 relative humidity Temperature 5 to 40 C Storage and transport Temperature range 20 to 60 C 4 to 140 F Atmospheric pressure 450 to 1100 hPa Humidity 15 to 95 relative humidity Performance Standard Volumetric Mean Diameter VMD Nebulizer Unit Flow rate 0 3 mL min Particle size VMD between 4 0 µm a...

Страница 14: ...Aeroneb Lab Nebulizer System Instruction Manual 14 Manufacturer Aerogen Ireland Limited Galway Business Park Dangan Galway Ireland Customer Service Telephone 353 91 502550 ...

Отзывы: