Aeroneb
®
Lab Nebulizer System Instruction Manual
12
Troubleshooting
Table 3: Aeroneb
®
Lab Troubleshooting
If this happens:
It could mean:
Try this:
No liquid in nebulizer unit.
Refill liquid through filler cap in the
nebulizer unit
(
see
Adding Liquid
).
Nebulizer unit has not been properly
cleaned.
Clean nebulizer unit (see
Maintenance
).
The indicator lights, but
aerosol is not visible.
It may be time to replace the nebulizer
unit.
Refer to Aeroneb
®
Lab parts list.
Indicator does not light.
There is no power to the system.
Verify that AC/DC adapter is
securely attached to control
module.
Nebulizer unit has not been properly
cleaned.
Clean nebulizer (see
Maintenance
).
Longer than expected
treatment time.
The nebulizer unit may need
replacement.
Refer to Aeroneb
®
Lab parts list.
If these suggestions do not correct the problem, discontinue use of any device that is, or appears to
be damaged or not operating properly and contact your local Aeroneb
®
Lab product representative.
Order Numbers
Table 4 lists the Aeroneb Lab order numbers (see Figure 1 for pictures)
Table 4: Aeroneb
®
Lab parts list.
Description Order
Number
Aeroneb
®
Lab Control Module (includes DC cable)
AG-AL7500
Aeroneb
®
Lab Control Module (includes AC/DC adapter)
AG-AL7000
Aeroneb
®
Lab Nebuliser Unit, Autoclavable, Standard (4.0 –
6.0 VMD) particle size.
AG-AL1000
Aeroneb
®
Lab Nebulizer Unit, Autoclavable, Small (2.5 – 4.0
VMD) particle size.
AG-AL1100
Aeroneb
®
Lab DC Power Cable
AG-AL2000
Control Module detachable cable (replaceable)
AG-AP1085
AC/DC Adapter
AG-AP1040-XX*
Filler Cap (pack of 5)
AG-AP1030
Aeroneb
®
Lab Instruction Manual
AG-AL1010
*Consult your local representative for the order code extension specific to your country and for pricing.