Clamp-on Ground Resistance Tester Model 6416
7. SPECIFICATIONS
7.1 REFERENCE CONDITIONS
Influencing Parameters
Reference Values
Ambient temperature
73 °F ± 5 °F (23 °C ± 3 °C)
Relative humidity
50 % RH ±10 %
Battery voltage
6 V ± 0.2 V
Magnetic field
< 40 A/m DC (no AC field)
Electric field
< 1 V/m
Operating position
Clamp horizontal
Position of the conductor in
the clamp
Centered
Measurement environment
No adjacent conductors carrying current within
10 cm
Proximity to magnetic mass
> 10 cm
Loop resistance
Non-inductive resistance
(20 Ω for the voltage measurement)
Measured current, sinusoidal
frequency
Frequency 50/60 Hz
Level of distortion < 0.5 %
Inferior current in loop
resistance measurement
Zero for the resistance and inductance
measurements
< 3.75 A for the voltage measurement
7.2 ELECTRICAL
7.2.1
Loop Resistance Measurement
Measurement Range
Resolution
Accuracy
0.010 Ω to 0.099 Ω
0.001 Ω
±1.5 % ±0.01 Ω
0.10 Ω to 0.99 Ω
0.01 Ω
±1.5 % ±0.02 Ω
1.0 t Ω o 49.9 Ω
0.1 Ω
±1.5 % ±0.1 Ω
50.0 Ω to 99.5 Ω
0.5 Ω
±2 % ±0.5 Ω
100 Ω to 199 Ω
1 Ω
±3 % ±1 Ω
200 Ω to 395 Ω
5 Ω
±5 % ±5 Ω
400 Ω to 590 Ω
10 Ω
±10 % ±10 Ω
600 Ω to 1150 Ω
50 Ω
20 % approximately
1200 Ω to 1500 Ω
50 Ω
25 % approximately
Alarm: range of threshold from 1
Ω
to 199
Ω
.