![AEMC instruments 6416 Скачать руководство пользователя страница 16](http://html1.mh-extra.com/html/aemc-instruments/6416/6416_user-manual_2861034016.webp)
16
Clamp-on Ground Resistance Tester Model 6416
2.8 BUTTONS FUCTIONS
Button
Description
▲
•
When the rotary switch is set to Ω+A or A: Increases the
brightness of the display, making it easier to read the display
in an environment with strong background illumination.
Highlighting activated for 30 seconds.
• When the rotary switch is set to SET-UP or MR: Serves as
the ▲ arrow when browsing in the menus and values. The
brightness of the display does not change when device is set
to SET-UP or MR.
MEM
▼
•
When the rotary switch is set to Ω+A or A: Records the
measured value. All of the data is recorded in the Standard or
Advanced mode.
• When the rotary switch is set to SET-UP or MR: Serves as
the ▼ arrow when browsing in the menus and values.
►
•
When the rotary switch is set to Ω+A (Advanced Mode):
Short Press:
Switches the display through the following three
modes:
- Display of the impedance recalculated at the selected
frequency.
- Display of the contact voltage (product Z*I).
- Display of R and L.
Long Press:
Enables or disables the audible alarms.
• When the rotary switch is set to SET-UP:
- Validation when browsing in the menus and values.
• When the rotary switch is set to MR (Advanced Mode):
- Switches the display through the measurement screens
and the measurement date/time.
HOLD
&
AUTO-HOLD
•
When the rotary switch is set to Ω+A:
- Freezes the displayed measurement for as long as the
HOLD
button is pressed. The NOISE, clamp open
(
), and alarm exceedence ( ) icons are visible,
if they were active.
AUTO-HOLD
Function:
If the AUTO-HOLD mode was enabled during instrument
SET-UP (see § 4.4, menu #11), opening the clamp acts as
the identical state to the HOLD mode for as long as the clamp
is open. The benefit of this function is that it makes it easy to
freeze the measurement with one hand, when access to the
HOLD button is difficult.
If the HOLD button is not pressed during this time, closing the
clamp automatically exits the AUTO-HOLD mode.