Use of the CellAED®
Do not attempt to operate the CellAED® unless thoroughly familiar with this manual and the functions of all controls, indicators and of
the electrode pads. Users must ensure that they have read and understand the user manual before use. Use the CellAED® exactly as
described in this manual.
The CellAED® has a differentiated use for infants weighing less than 10 kg (22 lbs). You must use Infant Mode for an infant weighing less
than 10 kg (22 lbs). Failure to do so may result in ineffective treatment.
When using the CellAED® in Infant Mode and removing the front electrode pad to begin chest compression CPR, take care to place the
electrode pad gel-side up when removed. Do not touch the exposed electrode or stick the pad to any object, as this may contaminate
the electroconductive gel and compromise the performance of the CellAED®.
Correct pad placement and good pad contact with the skin is critical.
Do not remove the electrode pads from the patient during or after rhythm recognition, during charging, or while the shock is being
delivered. The shocks delivered by the CellAED® can cause serious harm to user(s) or bystanders if the instructions are not followed.
Do not touch the pads to each other outside of deactivation protocol, as this may damage the electrodes. Do not stick the electrode
pad to any object.
Patient handling
The energy emitted by the CellAED® can be conducted through the patient or conductive matter touching the patient. Therefore:
• Do not touch the patient, electrode pads, metal objects such as a bedframe or stretcher in contact with the patient, or any other
material in contact with the patient while the shock is being delivered. These shocks can cause serious harm to user(s) or bystanders if
the instructions are not followed.
• Do not use the CellAED® when the patient is in or is near a pool of water or on any other conductive surface(s). Carefully remove the
patient if necessary. Ensure the patient is dry before using the CellAED®.
Pregnant women must be lain on their left side in the recovery position.
Содержание CellAED
Страница 24: ...SINGLE USE AED AUTOMATED EXTERNAL DEFIBRILLATOR PEEL SNAP STICK 2 24 FIGURE 3A CellAED outer box front...
Страница 25: ...25 FIGURE 3B CellAED outer box back QR codes...
Страница 27: ...27 FIGURE 4B CellAED protective case inside...
Страница 39: ...39 Continued page 40 4 2 3 1 FIGURE 10 Move patient into recovery position...
Страница 70: ......