For transmitters rated at a maximum output power
not listed above, the recommended separation
distance d in meters (m) can be determined using
the equation applicable to the frequency of the
transmitter, where P is the maximum output power
rating of the transmitter in watts (W) according to
the transmitter manufacturer.
Note 1:
At 80 MHz and 800 MHz, the separation
distance for the higher frequency range applies.
Note 2:
The ISM (industrial, scientific and medical)
bands between 150 kHz and 80 MHz are 6.765 MHz
to 6.795 MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz
to 27.283 MHz; and 40.66 MHz to 40.70 MHz.
Note 3:
An additional factor of 10/3 is used in
calculating the recommended separation distance
for transmitters in the ISM frequency bands between
150 kHz and 80 MHz and in the frequency range
80 MHz to 2.5 GHz to decrease the likelihood that
mobile/portable communications equipment could
cause interference if it is inadvertently brought into
patient areas.
Note 4:
These guidelines may not apply in all
situations. Electromagnetic propagation is affected
by absorption and reflection from structures,
objects, and people.
Содержание CellAED
Страница 24: ...SINGLE USE AED AUTOMATED EXTERNAL DEFIBRILLATOR PEEL SNAP STICK 2 24 FIGURE 3A CellAED outer box front...
Страница 25: ...25 FIGURE 3B CellAED outer box back QR codes...
Страница 27: ...27 FIGURE 4B CellAED protective case inside...
Страница 39: ...39 Continued page 40 4 2 3 1 FIGURE 10 Move patient into recovery position...
Страница 70: ......