59
IMMUNITY TEST
IEC 60601-1-2:2014/EN 60601-
1-2:2015
TEST LEVEL
COMPLIANCE LEVEL
ELECTROMAGNETIC ENVIRONMENT
Electrostatic Discharge (ESD) IEC
61000-4-2:2008, EN61000-4-2:2009
±
2kV ,
±
4kV,
±
6kV for Direct &
Indirect Contact
±
2kV,
±
4kV,
±
8Kv
for Air Discharge
±
2kV ,
±
4kV,
±
6kV for Direct &
Indirect Contact
±
2kV,
±
4kV,
±
8Kv
for Air Discharge
There are no special requirements with respect to
electrostatic discharge.
Power-frequency magnetic field IEC
IEC 61000-4-8:2009/EN 61000-4-
8:2009
30 A/m
50Hz
30 A/m
50Hz
Power-frequency magnetic fields should not be
greater than magnitudes which are typical of
commercial or hospital environments. There are
no special requirements for non-commercial/non-
hospital environments.
The CellAED® is intended for use in the electromagnetic environment specified below. The users must
ensure that it is used in such an environment.
10.5.2 ELECTROMAGNETIC IMMUNITY (IEC 60601-1-2:2014/EN 60601-1-2:2015 TABLE 202)
(Continued page 60)