7.
SPECIFICATIONS
7-1
7.
SPECIFICATIONS
7.1
R3755A Specifications
This chapter describes about the function of the R3755A and the performance/specification together.
1.
Measurement Function
Measurement channels
4 channels
Measurement parameter
A/R
Measurement format
Log / Linear amplitude, Phase, complex parameter real
and imaginary numbers,
Z, R, X (When measuring impedance conversion)
Y, G, B (When measuring admittance conversion)
2.
Signal Source Section (25
°
C
±
5
°
C, and calibration period: 1 year)
Frequency characteristics
Range
Resolution
Accuracy
10 kHz to 300 MHz
1 mHz
±
20 ppm
±
1 ppm (Opt20)
Output characteristics (PI circuit : ON)
Range
0 dBm to -30 dBm (10 kHz to 1 MHz)
+18 dBm to -43 dBm (1 MHz to 300 MHz)
Resolution
0.1 dB
Accuracy
±
2.0 dB (0 dBm, 10 MHz)
Linearity (10 MHz)
+18 dBm to -35 dBm
±
1.5 dB
-35 dBm to -43 dBm
±
2.5 dB
Flatness (At 0 dBm Output)
1 MHz to 100 MHz 5.0 dBp-p
100 MHz to 300 MHz 8.0 dBp-p
Sweep characteristics
Sweep parameter
Frequency
Range
Frequency sweep same as frequency characteristics
Range setting
Start / Stop or Center / Span
Sweep type
Any specified segment sweep
(Frequency, Output level, RBW, Point, Settling setting)
Sweep time
Max. 0.05 ms / Point (RBW 15 kHz)
Measurement points (Segments)
Max. 1601 points (segments)
Sweep trigger
Continuous, Single, or External trigger
Sweep mode
Single channel sweep
Output format
Output
Connector
Single
SMA (Female), 50
Ω