A.3.1
Descriptions of Window Components
A-21
Ele Corr
Electrical length correction:
Sets ON/OFF for electrical length correction.
Electrical length correction Time:
Electrical length correction length:
Sets the electrical length correction value by time or length.
(The units are psec and mm.)
Port Extension:
Sets ON/OFF for the port extension function.
Port Extension Time:
Port Extension length:
Sets the port extension function by time or length.
(The units are psec and mm.)
Figure A-26 Ele Corr
Extension of measurement reference plane
Assuming that a completely lossless cable has been added, electrical length for the addition is corrected.
By correcting the phase shift for the extension, phase characteristics for the sample can be obtained.
•
Electrical length correction (Electrical Delay)
The electrical length set to measurement data is corrected. Any measurement port can be used.
This function can be used when measuring cable electrical length.
The electrical length can be set for each trace. The electrical length can be set by time or length. Sets
ON/OFF for electrical length correction using the Electrical length correction check box.
•
Phase offset
A constant phase value is added as offset regardless of frequency. The phase offset can be set for each
trace.