59
APPENDIX B. RACK MOUNTING
The Model 273A can be mounted in a standard 19 inch relay rack by means of the Princeton
Applied Research 1715-0206 Rack Mounting Kit. This kit consists of two special side panels with
integral mounting flanges and four screws. To install the kit, proceed as follows.
1. Remove the side panels from the instrument. Each side panel is secured by four screws. DO
NOT DISCARD THESE SCREWS.
2. Mount the side panels supplied with the kit. The flanges should go toward the front and turn
outwards. Secure each panel with SIX screws. You will need the four screws supplied with
the kit as well as the eight screws that became available when the original side panels were
removed.
This completes the procedure. The 273A can then be mounted in any standard rack and secured
with standard rack-mounting screws (not supplied). Because of the Model 273A's considerable
weight, we strongly recommend that it be supported to prevent excessive stressing of the front
panel. One way of providing the necessary support is to set the Model 273A on a rack-mounted
shelf or drawer.