Chapter 3—Characteristics
19
Fig. 4. Simplified Block Diagram of Model 273A/92
Electrochemical Impedance Interface Option.
3.2.10. 273A/92 Electrochemical Impedance Interface Option
The 273A/92 Electrochemical Impedance Interface option (previously called the AC Interface
option) is intended primarily for use with Princeton Applied Research electrochemical
impedance systems, but can also be used for ac voltammetry applications. This option is
designed to allow superposition of an externally generated ac excitation signal on the dc signal
generated by the Model 273A.
The externally generated ac signal is input at the AC INPUT connector and is combined with the
signal generated by the Model 273A before it is applied to the cell (see Fig. 4). The /92 option
board has three output connectors (AC I OUTPUT, AC E OUTPUT, and MULTIPLEXED
OUTPUT. Software commands control both the mode of output (I, E, or multiplexed) as well as
the type of signal applied (full ac + dc signal or ac signal only). Multiplexing is accomplished by
alternating the operand of the MIE command between "1" and "2" on successive iterations. The
Model 273A Remote-Programming Command Handbook describes the computer commands
used with the Model 273A/92 option.
Содержание 273A
Страница 1: ...0RGHO 3RWHQWLRVWDW DOYDQRVWDW 8VHU V 0DQXDO 8VHU V 0DQXDO 3ULQWHG LQ 86...
Страница 6: ...vi Model 273A Potentiostat Galvanostat User s Manual...
Страница 10: ...4 Model 273A Potentiostat Galvanostat User s Manual...
Страница 23: ...Chapter 3 Characteristics 17 Fig 3 Model 273A Simplified Block Diagram...
Страница 26: ...20 Model 273A Potentiostat Galvanostat User s Manual...
Страница 66: ...60 Model 273A Potentiostat Galvanostat User s Manual...
Страница 70: ...64 Model 273A Potentiostat Galvanostat User s Manual...