50
Model 273A Potentiostat/Galvanostat User’s Manual
One of the two terminals of each banana plug pair has a small tab and is labeled "ground." This
"ground" terminal connects to the recorder terminal (X or Y as appropriate) marked “-" (Houston)
or LO (Hewlett Packard). The recorder terminal labeled GROUND on Houston recorders is not
used in operation with the Model 273A.
A relay contact closure inside the Model 273A is brought out to the pen drop leads. These leads
float with respect to ground. Many recorders have a pen-drop circuit configured such that the pen
drops when the closure occurs.
5.3.3. Auxiliary Interface
Various miscellaneous functions are provided by this connector, including.
1. Ground.
2. +5 V (20 mA maximum; sufficient for pullups).
3. Signals for interfacing to the Model 303A SMDE (PURGE, STIR, and DISLODGE).
4. Various inputs and outputs controlled by commands applied via the GPIB or RS-232C ports.
See Appendix A for more detailed information.
5.3.4. Power Amp Monitor
The output of a 10:1 attenuator connected to the Counter Electrode potential is provided at this
connector. The output impedance is 1 k
.
5.3.5. Aux A/D Input
This high-impedance (>10
) input provides an external input to the A/D Converter. This input
9
allows users to track any experimental variable that has an analog voltage output in the range of
± 10 V. For example, with the appropriate ancillary instrumentation (not available from Princeton
Applied Research), a user might track temperature, pH, optical absorbance, or mechanical
displacement. One specific use for this input is in making galvanic corrosion measurements with
the Model 332 Corrosion Measurement System. Note that there is no way of reading the applied
level from the front panel. A reading can be initiated only via commands applied to the
rear-panel RS232C or GPIB ports.
5.3.6. Electrometer Monitor
The output of the differential electrometer is available at this connector. This output is NOT
corrected for R compensation (unlike the front-panel E MONITOR Output). The output
U
impedance is 1 k
. Note that this potential is inverted in sign with respect to the applied cell
potential.
5.3.7. AC I Output
This connector, used only in units equipped with the 273A/92 Electrochemical Impedance
Interface option, outputs the ac current signal from the test cell. The signal is controlled by
commands from an external computer via the GPIB or RS-232C port (refer to the Model 273A
Remote Programming Command Handbook).
5.3.8. AC E Output
This connector, used only in units equipped with the 273A/92 Electrochemical Impedance
Interface option, outputs the ac voltage signal from the test cell. The signal is controlled by
commands from an external computer via the GPIB or RS-232C port (refer to the Model 273A
Remote-Programming Command Handbook).
Содержание 273A
Страница 1: ...0RGHO 3RWHQWLRVWDW DOYDQRVWDW 8VHU V 0DQXDO 8VHU V 0DQXDO 3ULQWHG LQ 86...
Страница 6: ...vi Model 273A Potentiostat Galvanostat User s Manual...
Страница 10: ...4 Model 273A Potentiostat Galvanostat User s Manual...
Страница 23: ...Chapter 3 Characteristics 17 Fig 3 Model 273A Simplified Block Diagram...
Страница 26: ...20 Model 273A Potentiostat Galvanostat User s Manual...
Страница 66: ...60 Model 273A Potentiostat Galvanostat User s Manual...
Страница 70: ...64 Model 273A Potentiostat Galvanostat User s Manual...