Calibrating the PCS Instrument
From the
!
Home
screen, click
Due
instruments if the count is not zero and
then skip to step 3. Otherwise, click the
Shortcut for
Calibrate Instrument.
If calibrating from the Shortcut, select the
instrument from the instrument list and a
schedule if that is the calibration type.
Click
Start Calibration.
Follow the on-screen instructions to
scan and insert
PCS Instrument
Calibrator Kit
vials A through D.
ArtelWare requires a Passing instrument calibration before performing pipette calibrations or operator trainings.
Artel recommends creating an Instrument Calibration schedule for every 30 days.
Note
: The first time calibrating the PCS Instrument, ArtelWare requests the
ArtelWare Activation
code located inside the PCS Calibrator kit box. To use the Activation code before calibrating the
instrument, click ‘
Scan Barcode
’ on the ArtelWare
!
Home
screen and then follow the prompts in ArtelWare. This activates ArtelWare until the expiration of the date of the Calibration kit
.
1
2
3
9
1
2
3
4
4
If calibrating from the Shortcut, select the
Calibration Type and click
OK
.