Uploading External Calibration Records
Upload pipette calibrations completed outside of ArtelWare by selecting the
Ad Hoc Pipette
screen under the
%
Calibration
menu. or
Ad hoc Calibration
from the Calibration Type screen after clicking the
Calibrate Pipette
shortcut on the
!
Home
screen
Choose External Pipette
Calibration for the
Calibration
Type
and then select an
appropriate
Calibration Reason
before clicking
Start Calibration
.
Enter the Calibration Date and
the Calibration Result.
Click
Browse
to select a
calibration document exported
from Artel Data Manager, Artel
Pipette Tracker, or from a third-
party service or instrument.
Optionally, use
+
Create new
Summary
to enter Summary
Statistics for the Calibration
Report.
Click
Upload External
Calibration
to save the record
in ArtelWare.