Performing Unscheduled Volume Measurements
14
To measure a volume dispensed by a device without requiring a calibration plan, use the Measure Volume option from the
!
Home
screen. Measurements
made with this option do not get recorded in the system database, no report is generated, and results are discarded after ending the measurement cycle.
1
2
3
From the
!
Home
screen Shortcuts bar,
click
Calibrate Pipette
. Choose
PCS Volume
Quick Check
for the Calibration Type.
Enter the expected target volume in µL, the
number of replicates to dispense, and select to
display errors as relative or absolute.
Select the instrument to use, click
OK
and then
follow the on-screen instructions for
completing the volume measurement session
.
Ad hoc Calibration
(alternative method)
Choose this Calibration Type to use a Calibration
Plan without affecting due dates for the pipette.
However, the calibration data
will be
saved in the
database for future reference
.
*