22
Human NIBP Nano
Owner’s Guide
INCORRECT CUFF WRAPPING
Wrong
– Cuff is rotated 180°.
Wrong
– Cables point in the
wrong direction.
The photo-
plethysmograph
is not positioned
in front of the
arteries.
The photo-
plethysmograph
is not positioned
in front of the
arteries.
bone
LED
photodiode
arteries
Finger cross-section
CORRECT CUFF WRAPPING
Wrong
– Cuff is shifted
proximally.
Wrong
– Cuff is rotated.
The conical shape
of the cuff does
not match the
fi nger..
The middle phalanx
is the preferred
position.
Figure 2–10 : Cuff wrapping
instructions
1. Point cable and
tube towards
wrist.
2. Center cuff
between joints.
3. Center LED and
photodiode
symmetrically.
4. Wrap cuff tightly.
5. Place hand on
side, thumb up.