18
Human NIBP Nano
Owner’s Guide
Using Finger Cuffs
Proper finger cuff application is critical to successful finger arterial
pressure measurement with the Human NIBP Nano Interface. Therefore,
read the following sections about finger cuff handling, selection and
application with special attention!
Warnings on cuff handling
Take into account the following notes to prevent finger cuff damage:
•
Don’t apply air pressure to a finger cuff when it is not wrapped
around a finger (or any other solid object!). This will damage the
finger cuff .
•
Don’t bend finger cuff s outwards into a flat shape since this may
damage the bonding and the electrical shielding. Finger cuff s are
preformed around a conical mandrel during manufacturing.
Figure 2–6
Final placement
of the Height
Correction Unit
(HCU)
1 – Place reference end at heart level.
2 – Attach the HCU transducer to the fi nger cuff.
1
1
2
C1
HCU
C2
C1
HCU
C2