25
Human NIBP Nano
Owner’s Guide
•
Active Cuff :
This indicates which finger cuff is active during that
part of the recording.
•
Cuff Countdown:
This channel shows the countdown of the
number of seconds until finger cuff switching occurs.
•
AutoCal Quality:
This is an internal measurement of the quality of
the last AutoCal performed.
•
AutoCal Countdown:
This channel shows the countdown of the
number of heart beats until the next AutoCal is performed.
Human NIBP Settings
This section describes the options available in the Human NIBP Settings
dialog. Go to
Setup > NIBP Settings...
to open this dialog.
Auto calibration enabled
Auto calibration is based on a patented model for blood pressure
calibration (see details of the PhysioCal algorithm on page 48). Auto
calibration or ‘AutoCal’ is on by default. It improves accuracy in the arterial
blood pressure measurement by providing an ongoing calibration, every
Figure 2–12
Opening the Human NIBP
Settings dialog.