Electromagnetic Compatibility
The Human NIBP Nano Interface has been tested to comply with the requirements
of EN60601-1-2 + A1: medical electrical equipment; EMC requirements and tests.
The device also conforms to the following safety standards: CORR.2 (2007) and
Canadian/US diff erences according to CAN/CSA-C22.2 No. 60601-1:08, and ANSI/
AAMI ES60601-1:2005.
Emissions
•
The Human NIBP Nano Interface uses RF energy for its internal function
only. RF emissions are very low and are not likely to cause any interference
in nearby electronic equipment.
•
The Human NIBP Nano Interface is suitable for use in all establishments,
including domestic establishments and those directly connected to the
public low-voltage power supply network that supplies buildings used for
domestic purposes.
Immunity
•
Mains power quality should be that of a typical commercial or hospital
environment. If the user of the devices requires continued operation
during power mains interruptions, it is recommended that the devices be
powered from an uninterruptible power supply or a battery.
•
Power frequency magnetic fields should be at levels characteristic of a
typical location in a typical commercial or hospital environment.
•
Floors should be wood, concrete or ceramic tile. If floors are covered with
synthetic material, the relative humidity should be at least 30%.