44
Table 19 Pressure delta setting
Project
Effective value
Description
Enable
Enable / disable
Enable or disable the differential pressure module. When the
differential pressure module is enabled, the differential pressure
module channels will be displayed in the related functions
channel list.
Resolution
4/5/6
The resolution of differential pressure
Range
Numbers
Differential module measuring range
Calculate type
A-B or B-A
For the composite mode of differential pressure module, A and B
represent external pressure module A and external pressure
module B respectively
Real time data
------
or Real time differential pressure value
The real-time differential pressure value is displayed only when
the differential pressure module is enabled and both external
pressure modules A and B are online. Otherwise, the real-time
differential pressure value is displayed as ---------
Zero offset
zero offset when zeroing
Zero offset when differential pressure module zeroing
External module A
Module real time pressure values and units
Click to enter to view the module information
External module B
Module real time pressure values and units
Click to enter to view the module information
The following functions are also supported in the DP module management.
Zero:
when enabling the DP module and both external A and B are online, users can perform the zero, the zero offset will
be saved and used for calculating the differential pressure.
Cancel zero:
cancel the zero offset when performing the zero operation, set the zero offset to 0.
Содержание 273E Series
Страница 1: ...1 273Ex Handheld Pressure Calibrator...
Страница 2: ......
Страница 3: ......
Страница 4: ...ADT273Ex Handheld Pressure Calibrator User Manual Version 2302V02 Additel Corporation...
Страница 6: ......
Страница 20: ...10 1 3 Basic Structure Figure 1 Basic Structure...
Страница 41: ...31 Figure 16 Sensor trim Figure Figure 17 Process variable selection...
Страница 56: ...46...