40
Output range
Numbers
Output range of DUT
Transfer
function
Linear / square root
Transfer function of the DUT
Accuracy
0.06%, 0.1%, 0.16%, 0.25%, 0.4%, 0.6%,
1%, 1.6%, 2.5%, 4%, custom
Accuracy of DUT. If choose custom, the input number means the
accuracy level of the DUT. For example: input 1.5 for level of 1.5,
the range is (0.001~100)
Process
variable
Process variable/ output current/ percentage/
loop current
Process variable of HART transmitter(only available for HART
transmitter)
3. Start the task
•
Please refer to the “Start the Task” in Section 6.1.
•
The calibrator will automatically use the measured value of the voltage measurement, current measurement
or the HART measurement according to the DUT output type set by the user, without manual input.
4.Finish the task
Refer to “Finish the Task” in Section 6.1.
6.3 Pressure switch
1. Connect to DUT
•
Connect the internal pressure module. If use external pressure module, connect them to LEMO port A or B.
•
As shown in the Figure 6, connect the DUT output with calibrator’s electric module.
2. Create the task
Create the corresponding DUT tasks as described above, and enter the necessary information.