5
Input parameters, simple connection
Connection
Function
Ui/V
Ii/mA
Pi/mW
Ci/nF
Li/mH
J1
、
J3
HART (external power and external
resistance)
30
100
750
10
0
J2
、
J3
Current measurement
30
100
750
1
0
J4
、
J3
Voltage, frequency and switch
measurement
30
100
750
1
0
J1
、
J2
Loop Circuit Transmitter Current
Measuring / HART(Internal Power and
Resistance)
30
100
750
10
0
Output parameters, simple connection
Connection
Function
Uo/V
Io/mA
Po/mW
Co/µF
Lo/mH
J1
、
J3
HART (external power and external
resistance)
25.2
79.16
499
0.107
9
J2
、
J3
Current measurement
5
0.85
1.1
87
100
J4
、
J3
Voltage, frequency and switch
measurement
5
0.85
1.1
87
100
J1
、
J2
Loop Circuit Transmitter Current
Measuring / HART(Internal Power and
Resistance)
25.2
79.16
499
0.107
9
Intended use
•
The equipment to be used in Zone 0 hazardous area with gas group IIC.
•
The equipment is used in oil and gas platforms, oil refineries, chemical and petrochemical plants,
pharmaceutical industry, energy and gas processing industries.
Содержание 273E Series
Страница 1: ...1 273Ex Handheld Pressure Calibrator...
Страница 2: ......
Страница 3: ......
Страница 4: ...ADT273Ex Handheld Pressure Calibrator User Manual Version 2302V02 Additel Corporation...
Страница 6: ......
Страница 20: ...10 1 3 Basic Structure Figure 1 Basic Structure...
Страница 41: ...31 Figure 16 Sensor trim Figure Figure 17 Process variable selection...
Страница 56: ...46...