7352 Series Digital Multimeter Operation Manual
6.6.4 SCPI Command Reference
6-38
Calculation
dB/dBm
calculation
:CALCulate:DB {DB|DBM|OFF}
[,(@<ch>)]
DB: dB calculation ON
For the voltage and current measurement functions,
ON can be set.
DBM: dBm calculation ON
For the voltage measurement function, ON can be
set.
OFF: dB/dBm calculation OFF
:CALCulate:DB?[(@<ch>)]
Reply: DB, DBM, or OFF
:CALCulate:DB:D
{<real>|MAXimum|MINimum|
MEASurement}[,(@<ch>)]
Constant D
<real>: 0.00001E-9 to 999999.E+6
(1)
:CALCulate:DB:D?
[{MAXimum|MINimum}][,(@<ch>)] *3
Reply:
d.dddddE
dd
MAX•MIN
calculation
:CALCulate:AVERage
{ON,1|OFF,0}[,(@<ch>)]
ON, 1: ON
OFF, 0: OFF
:CALCulate:AVERage?[(@<ch>)]
Reply: 1 or 0
:CALCulate:AVERage:MAXimum?
[(@<ch>)]
MAX value reading
Reply:
d.dddddE
dd
:CALCulate:AVERage:MINimum?
[(@<ch>)]
MIN value reading
Reply:
d.dddddE
dd
:CALCulate:AVERage:AVERage?
[(@<ch>)]
AVE value reading
Reply:
d.dddddE
dd
:CALCulate:AVERage:COUNt?[(@<ch>)]
Measurement count reading
Reply:
d.dddddE
dd
Comparator
calculation
:CALCulate:LIMit {ON,1|OFF,0}
ON, 1: ON
OFF, 0: OFF
:CALCulate:LIMit?
Reply: 1 or 0
:CALCulate:LIMit:UPPer
{<real>|MAXimum|MINimum|
MEASurement}
HIGH constant
<real>: 0.00001E-9 to 999999.E+6
Setting resolution:0.000001E-9
:CALCulate:LIMit:LOWer
{<real>|MAXimum|MINimum
|MEASurement}
LOW constant
<real>: 0.00001E-9 to 999999.E+6
Setting resolution:0.000001E-9
:CALCulate:LIMit:UPPer? [
{MAXimum|MINimum}]
Reply:
d.dddddE
dd
:CALCulate:LIMit:LOWer?
[{MAXimum|MINimum}]
*3: If omitting [{MAXimum|MINimum}], do not set “,” followed by (@<ch>).
Item
SCPI command
Description
Initial