ADE- 2022-C
Page 31 of 50
Variable
Variable
Variable
Variable
Value
Value
Value
Value
Units
Units
Units
Units
Description
Description
Description
Description
Digital imputs
Digital imputs
Digital imputs
Digital imputs
d1P
OP
Drawer switch polarity
Odc
F-C
Open drawer control
dOA
5
Open drawer alarm delay
dLc
Yes
Stop count down of running cycle
rrd
No
Regulation restart after door open alarm
d2F
EAL
Second digital input function
d2P
cL
Second digital input polarity
did
15
Time delay for digital input alarm
Auxiliary relay configuration
Auxiliary relay configuration
Auxiliary relay configuration
Auxiliary relay configuration
oA1
C2
First configurable relay function
oA2
ALL
Second configurable relay function
oA3
Lig
Second configurable relay function
Second relay management
Second relay management
Second relay management
Second relay management
2CH
C1
Compressor setting during the holding
OAt
0
Second compressor switching on delay
OAS
0.1
Set point for second compressor
OAH
0.1
Differential for second compressor
OAi
rP
Probe selection for second compressor
Auxiliary relay management
Auxiliary relay management
Auxiliary relay management
Auxiliary relay management
OSt
0
Auxiliary output timer
OSS
-30
Set point for auxiliary output
OSH
0.1
Differential for auxiliary output
OSi
rP
Probe selection for auxiliary output
Defrost
Defrost
Defrost
Defrost
tdF
rE
Defrost type
IdF
8
Interval between defrost cycles
dtE
8
Defrost termination temperature
MdF
25
Maximum length for defrost
dFd
DEF
Temperature displayed during defrost
Fdt
6
Drip time
dAd
30
Defrost display time out
Fans
Fans
Fans
Fans
FnC
O-n
Fans operating mode
FSt
20
Fan stop temperature
AFH
1
Differential for stop temperature and
alarm
Fnd
2
Fan delay after defrost