8.0 SERVICE
PARAMETERS
8.1 USING “0000” TO ENTER THE CALIBRATION PARAMETER
Calibration can be enabled by placing a jumper into the pins on the main PCB (marked as
position J4) through a hole provided on the base of the scale. After the jumper has been
placed in the position shown, apply power to the scale.
Turn the scale upside down.
Remove the seal covering the hole to access
the PCB.
Place a jumper through the hole into the pins. If
a jumper is not available, short the pins.
Turn on the scale using the switch on the base.
“
P
n
”
When
“
P
n
”
is displayed.
Enter
“
0 0 0 0
” and p
ress
[Tare]
“
U
n
LoA
d
”
Empty the pan by removing the load, if there is
any and press
[Tare]
“
L
o
A
d
” “
6
” “
K
i
L
o
S
”
Load the requested calibration weight and press
[Tare]
“
SPA
n
” “
PASS
”
If Calibration is complete,
“
SPA
n
PASS
”
will be
displayed. Remove the calibration weight.
Or, “
SPA
n
” “
FA
i
LE
d
”
This means calibration has failed. Remove the
calibration weight and repeat the process.
“
JP O
n
”
Remove the jumper or shorting of the pins
whichever is used. The scale will return to
normal weighing.
© Adam Equipment Company 2007