© Adam Equipment Company 2008
14
F6
init
ZEro
Initial Zero Range
F7
rE
ZEro
Re-Zero range
F8
SCSivE
tArE
Successive Tare Enable
F9
Ad
Count
Display the A/D counts
F10
round
Rounding off the Total Price
F11
ZEro
Select the zero mode
F12
Lvd
mod
Low voltage detection
F13
CAL
batt
Set calibration of battery voltage measurements
F14
ACLr
Not active,
F15
Cal Cnt
Enable or disable the Calibration Counting function
8.2.1
F1 -
CAL
To advance to the next parameter, press the
[0]
key or to enter the calibration parameter
press the
[Tare]
key when “
F1 CAL
” is displayed.
The display will instruct you to remove any weight from the scale, “
UnLoAd
”. Press
[Tare]
.
The display will tell you to add weight to the scale: “
LoAd
” “
6
” “
kiloS
”
Add the weight shown, wait for stability then press the
[Tare]
key.
The display will show “
SPAn
” “
PASS
” if the calibration is OK. Or it will show “
SPAn
” “
FAIL
” if
calibration could not be completed correctly.
The display will then show “
JP on
” indicating the jumper is still in place if it was used to
gain access. Remove the weight. Switch off the scale and switch it on to continue with
the other Service parameters.
8.2.2
F2
–100G
To enter this parameter, press the
[Tare]
key when “
F2 100G
” is shown.
The display will show “
PriCE
“ “
on
” or “
PriCE
“ “
oFF
” .
Содержание WBZ 15