© Adam Equipment Company 2008
20
8.2.14
F14
–AUTO CLEAR
This parameter enables or disables the auto-clear function.
To enter this parameter, press the
[Z/T]
key when “
F14
” is shown.
The display will show if the function is set to on or off.
Press the
[0]
key to change the setting.
Press
[Z/T]
to accept the displayed value and return to the parameter.
Press the
[0]
key to advance to the next parameter.
NOTE 1:
Only the “off” option is allowed for the scales to be used in South Africa. This will
disable the “Auto Clear” function on the
[3]
key and the auto-clear function will be
permanently on.
8.2.15
FFFF15
15
15
15
– CALIBRATION and PARAMETER COUNTING ENABLED
This parameter allows the scale to show the stored values for the number of times the
scale has been calibrated or parameters set. The Counting function is either set to On or
Off. If on the user has access to calibration and parameter setting without the need to
open the scale. See section 7.3. If set to Off the scale will not show the counter values
and the only way the user can access them is to use the jumper on the PCB as described
in section 8.1.
To enter this parameter, press the
[Tare]
key when “
F15
” is shown.
The display will show if the function is set to on or off.
Press the
[0]
key to change the setting.
Press
[Tare]
to accept the displayed value and return to the parameter.
Press the
[0]
key to advance to the first parameter.
To return to normal weighing, switch off the scale, remove the jumper and switch it on.
NOTE : AFTER SETTING PARAMETERS OR PERFORMING CALIBRATION
Содержание WBZ 15
Страница 3: ... Adam Equipment Company 2008 2 ...
Страница 24: ... Adam Equipment Company 2008 23 9 0 WIRING DIAGRAM ...
Страница 25: ... Adam Equipment Company 2008 24 10 0 MECHANICAL ASSEMBLY ...
Страница 26: ... Adam Equipment Company 2008 25 ...
Страница 28: ... Adam Equipment Company 2008 27 ...