3.0 RS-232 SPECIFICATION
The RS-232 is not available in WBW scales.
4.0 CALIBRATION
The WBW-M scales should not be calibrated by the user. Calibration of the scales may make it illegal to use
the scales. Contact your local metrology standards office for further assistance.
The WBW non-approved scales are calibrated using metric weights when the weighing unit selected is either
kilograms or grams and using pound masses when the weighing unit selected is either pounds, ounces or
pounds:ounces.
To start the calibration turn the scale off and then turn it on again. Press
[Tare]
during the self-test. Scale
will show
“P- - - - “
. Enter code number
“0000”
using the numeric entry method (see section2.1) and press
[Tare]
. This will take you directly to the calibration section.Display will show
"UnLoAd"
.
Remove all weight from the pan and then press the
[Tare]
key when the scale is stable. After the Zero point
is set, the display will show
“Ld xx”
. Place the suggested calibration mass on the pan. It is best to use a
weight close to the full capacity of the scale. If the mass is different from the displayed value, enter the value
of the mass in whole numbers using the numeric entry method (see section 2.1). The kg or the lb symbol will
be on to show the active unit.
Press the
[Tare]
key when the stable indicator is on. The scale will calibrate to the mass and then return to
weighing. Remove the calibration weight as soon as calibration is complete.