© Adam Equipment Co Ltd 2018
18
1) Press the
[Tare]
↵
key to accept the chosen format and then enter the date in this format.
2) Press the
[Tare]
↵
key to accept the date.
An error code will be shown if the time (Err 1) or the date (Err 2) is not the permissible values.
For example, 34
th
day of a month is an invalid entry.
5.0 ERROR CODES
During the initial power-on testing or during operation, the scale may show an error code in
which case consult the table below. If the error persists, contact your dealer for further support.
ERROR
DESCRIPTION
POSSIBLE CAUSES
Err 1
Time input error.
Tried to set an illegal time, i.e.
26 hours
Err 2
Date input error
Tried to set an illegal date, i.e.
36
th
day
Err 4
Initial Zero is greater than
allowed (typically 4% of
the maximum capacity)
when power is turned on
or when
[Zero]
is pressed,
Weight is on the pan when
turning the scale on.
Excessive weight on the pan
when zeroing scale. Improper
calibration of the scale.
Damaged load cell.
Damaged Electronics.
Err 6
A/D count is not correct
when turning the scale on.
Platform is not installed. Load
cell is damaged.
Electronics is damaged.
6.0 CALIBRATION
The SWZ scales are sealed to prevent unauthorised calibration.
Contact Adam Equipment or your supplier for more details.
WARNING
: CALIBRATION OF THE SCALES MAY MAKE IT ILLEGAL TO USE THE SCALES
FOR SALES OF GOODS. CONTACT YOUR TRAIDING STANDARDS OFFICE FOR
FURTHER ASSISTANCE.