Page 30 of 48
12.0 ERROR CHECKING
During weighing the balance is constantly checking to see if it is operating
within the limited parameters. The errors likely to occur are:
A/D counts below lowest allowed value
A/D counts above highest allowed value
A/D not operating
Maximum
capacity
exceeded
Other errors may be detected during special functions or operations. These
will be described in the section that applies.
Error messages and the reasons are:
Concerning A/D counts
ERROR ADc UL
A/D counts below a limit
ERROR ADc OL
A/D counts above a preset limit
Concerning calibration
ERROR StAb
Calibration could not be completed
because the results were not stable
ERROR LO or ERROR HI
Calibration constant not within 20% of old
calibration constant
Concerning weighing
ERROR LO
Weight display is below zero by >4%max
ERROR LOAd
Weight is above maximum plus 90d
© Adam Equipment Company 2007