LANGUAGE
English
French
German
Spanish
TIME
Set Time using the numeric entry method
(see section 7.1)
DATE DATE FORM
EUROPE (dd/mm/yy)
USA (mm/dd/yy)
Set Date using the numeric entry method
(see section 7.1)
INST ID
Enter a user number to identify this balance
BUZZER
On= Enable
OFF= Disable
BACKLIGHT
On
OFF
AUTO (default)
POWER DOWN
Set the time after which the unit will go into
Stand-by power settings, On=Enable,
OFF=Disable, If set to On-
the options are 1 to 9 minutes
FILTER
Set a value for the amount of filtering to be
done ranging from 1 to 10. A larger number
means more filtering and a slower
response.
STABILITY
Set a value to be used to determine
balance stability, set a value of 1, 2, 5 or
10d. A larger number corresponds to a
larger stable zone. Default is 5
AUTO ZERO
Can be set to On or OFF to enable the
auto-zero function. If set to On- select from
1, 2 or 5d
The sub-menu is entered by pressing
[Enter]
–
•
Use the
[Up]
and
[Down]
keys to increase or decrease the value for
setting. Press
[Enter]
to accept the setting and advance to the next
item in the menu
•
Press
[Mode]
to advance to setting of the next parameter or
[Esc]
to
return to normal weighing