13.5 CALIBRATION SETUP
This menu allows the user to set the calibration parameters.
•
Press
[Enter]
when “CAL SETUP” is displayed to select the calibration
parameters
•
The options for each parameter can be scrolled through by using the
[Up]
or
[Down]
key
ENABLE
NO =operator calibration is disabled
YES=operator calibration is enabled
CAL
REPORT
On = Enabled. Prints out Calibration
report after successful calibration
OFF = Disabled
TIME CAL
On = Enabled. Select time from 1 to 24
hours. Default setting is 4 hours
OFF = Disabled
TEMP CAL
On = Enabled. If set to On, set the
temperature variation from 0.2 to 4°C
OFF = Disabled
INT CAL
(Only on balances with
internal calibration masses)
YES = Use Internal calibration mass as
displayed or adjust for accurate Internal
Calibration (see section 13.5.1)
NO = Use external calibration mass
•
Press
[Mode]
to advance to setting of the next menu which is
“PASSCODES” or
[Esc]
to return to normal weighing.
13.5.1 INTERNAL
CALIBRATION MASS ADJUSTMENT
Balance model
Internal Calibration mass (approx.)
PGW 153i
100g
PGW 253i
240g
PGW 453i
240g
PGW 753i
520g
PGW 1502i
1400g
PGW 2502i
1400g
PGW 3502i
1400g
PGW 4502i
1400g
NOTE:
The value of the internal mass is not an exact number, such as 100g.
It must be set by comparing with an external mass. The exact value can be
set, provided this feature is enabled at the factory.