Basic Information (General)
2.1
Proper use
The balance you purchased is intended to determine the weighing value of material to be weighed.
It is intended to be used as a "non-automatic" balance, i.e. the material to be weighed is manually
and carefully placed in the centre of the weighing plate. As soon as a stable weighing value is
reached the weighing value can be read.
2.2
Improper Use
Do not use balance for dynamic weighing. In the event that small quantities are removed or added
to the material to be weighed, incorrect weighing results can be displayed due to the "stability
compensation" in the balance. (Example: Slowly draining fluids from a container on the balance.)
Do not leave permanent load on the weighing plate. This may damage the measuring system.
Impacts and overloading exceeding the stated maximum load (max) of the balance, minus a
possibly existing tare load, must be strictly avoided. Balance may be damage by this.
Never operate balance in explosive environment. The serial version is not explosion protected.
The structure of the balance may not be modified. This may lead to incorrect weighing results,
safety-related faults and destruction of the balance.
The balance may only be used according to the described conditions. Other areas of use must be
released by ACZET in writing.
2.3 Warranty
Warranty claims shall be voided in case
• Our conditions in the operation manual are ignored
• The appliance is used outside the described uses
• The appliance is modified or opened
• Mechanical damage and damage caused by media, liquids
Natural wear and tear
• The appliance is improperly set up or incorrectly electrically connected
• The measuring system is overloaded
5
Adjustment
Background illumination on
Contrastful display which can also be red
in the darkness.
Background illumination off
Battery saving
The background illumination will be
switched off automatically 10 sec after
having reached a stable weighing value.
Battery saving
Display
“EL on”
“EL off”
“EL AU”
Function
34
8.6.2
Display background illumination
In the menu display background illumination can be switched on or off.
To achieve this, follow the sequence of operations below.
MODE
MODE
Operation
Balance Display
MODE
•
Start balance by pressing
As long as the balance carries out a self test, press .
“F1 Unt” is displayed.
•
Press
“F2 bL” is displayed.
• Press
“EL on”
or
“EL off”
resp.
“EL AU”
is displayed
• Press to switch over between the three settings
• Press to activate the displayed setting.