In order to display the total weight of the samples, remove weight from the weighing plate, wait for
the zero display and press . The number of weighing actions and the total weight of the samples
will be displayed shortly.
With the further samples proceed as described above.
Press , „ACC 2“ is shortly displayed, then appears shortly
the total weight and then the weight of the second sample which now
is also stored.
The two following items limit the totalizing process:
1. Totalizing max. 99 times
2. Number of digits displayed
In order to delete the stored values, press and while the sum of the test
items is displayed, press , the balance will return to „0.0 g“.
27
“Pass” appears, the adjustment process has been finished
successfully.
• Remove adjusting weight during selftest
Wait until the balance is again in weighing mode.
5.8
Linearization
Linearity shows the greatest deviation of a weight display on the scale to the value of the respective
test weight according to plus and minus over the entire weighing range.
If linearity deviation is discovered during a testing instrument control, you can improve this by
means of linearization.
•
Carrying out linearization is restricted to specialist staff possessing well acquainted with
the workings of weighing scales.
•
The test weights to be used must be adapted to the weighing scale's specifications; see
chapter “testing instruments control”.
•
Observe stable environmental conditions. Stabilisation requires a certain warm-up time.
•
After successful linearization you will have to carry out calibration; see chapter “testing
instruments control”.
12