16
皇晶科技股份有限公司
Acute Technology Inc.
Copyright
2018
a long time, until the user manually stops capturing, or the set trigger
condition is met.
2. The total amount of data is less or equal to the total amount of device
memory.
3. The requirements for USB transfer rate or computer performance are low.
4. If you did not set a trigger, or you did but wish to retrieve the data before
the memory is full, you must manually press the stop button before the
data will be sent to the computer.
Device memory limitation
If unchecked, use the maximum memory of the device.
If checked, you can adjust the device memory usage. Set up lower memory
usage to shorten the processing time.
Data capture will continue until
pressing “Stop” (Wait for stop)
Even if the device memory is full, it continues capturing new data and
replaces the oldest ones. After you press
“Stop”, it sends back the newest
data stored in the device to the computer.
Data capture will continue until the trigger condition is met (Wait for
Trigger)
If the trigger condition is not set, there will be no Pre/Post Trigger
relationship and only the
Capturing
will be shown until the device memory
is full. If the trigger condition is set, data will be filled according to the
Trigger Position
settings. Data capture will continue until the trigger
condition is met or
Stop
is pressed, then it will stop by one of the three
actions:
Fill device memory then stop
Stop immediately
Wait for a few seconds then stop