1. Digital Channel Settings:
a. Choose the channel you want to measure, it sets threshold automatically, you
can adjust manually. 8 channels are a set of adjustable units, there are two
sets.
b.
(MSO1000, MSO2000 series only)
Provide Extra Hysteresis function, turn on
to reduce noise, turn off to increase sensitivity, replacing the past Schmitt
function.
c. The number of usable channels will vary depending on the trigger function
setting or sampling rate.
2. Analog Channel Settings:
For MSO1000, MSO2000
a. You can choose the channel you want to measure.
b. Input Sensitivity provides two different revolutions
(1). Voltage Range: +-10V, Minimum Scale: 5mV/Div.
(2). Voltage Range: +-20V, Minimum Scale: 10mV/Div.
For MSO3000
The number of available channels will be displayed in the analog setting column,
each channel is set separately.
ON / OFF
Enable/disable the channel.
Coupling
The input signal is DC/AC coupled. DC coupling does not
process the signal, while AC coupling removes the DC
level from the signal.
Bandwidth
The bandwidth limit can be set to Full bandwidth / 100
MHz / 20 MHz. Full bandwidth means that there is no