XI - ROUTINE
MAINTENANCE /
STERILIZATION
Important:
Except for the irrigation lines, no accessory is
delivered sterile.
During sterilization, items made of different
metals must not come into contact. Any contact
would establish galvanic couples and cause
local damage.
To avoid this happening, place the items in
individual sterilizable bags or in a sterilization
box.
To maintain the sterility or asepsis of the
accessories (contra-angle, micro-motor, micro-
motor cord, handpieces, etc.), store them in
sealed bags or containers suitable for medical
practice.
The maintenance and/or sterilization
instructions below must be applied before each
use of the device.
Avoid using cleaning and disinfection agents
containing flammable substances.
If this is not possible, make sure that all such
agent has evaporated and that there is no
combustible matter on the device or its
accessories before it is switched on.
Sterilized items must be allowed to dry and
cool to room temperature before they are
reused.
Before sterilization, check that your autoclave
is clean and that the water used is of adequate
quality.
After each sterilization cycle, remove the
items from the autoclave immediately in order
to minimize the risks of metal corrosion.
11. 1 IRRIGATION LINE MAINTENANCE
The irrigation lines supplied by SATELEC are
disposable and must always be discarded after
use.
Re-use of an irrigation line may result in
contamination of your patients.
Do not attempt to resterilize the irrigation lines.
Do not attempt to modify the irrigation lines.
11. 2 CONTRA-ANGLE MAINTENANCE
Refer to the manufacturer's instructions for your
contra-angle.
11. 3 ROTARY INSTRUMENT MAINTENANCE
Please refer to the manufacturer's instructions
for your instruments (burs, drills, etc.).
11. 4 DEVICE MAINTENANCE
Important:
Do not use an abrasive product to clean the
device.
Do not allow cleaning/disinfecting spray to
penetrate inside the device.
The device, the bracket and the multifunction
footswitch are not sterilizable.
The I
MPLANT
C
ENTER
control unit, the control
footswitch and the bracket must always be
cleaned and disinfected after each procedure,
using alcohol, disinfectant agents or disinfectant
wipes for dental practices, such as S
EPTOL
™ wipes.