Note
: While the contra-angle is being adjusted,
the maximum permitted speed and torque are
displayed automatically.
d) Speed adjustment
When the contra-angle has been selected or
entered, the cursor is placed below rpm (Fig. 5).
Press button . The cursor is placed under the
first figure.
Adjust the value of the first figure using the or
button (Fig. 3-7 and 9).
Press button . The cursor is placed under the
next figure.
Adjust the speed value figure by using the or
and buttons.
Confirm the speed value by pressing button .
Note
: The instrument speed can be adjusted
directly using the or button (Fig. 3-7 and 9),
including when the multifunction footswitch is
switched on.
To obtain a stable speed adjustment result, it is
recommended that the footswitch button be fully
pressed (Fig. 8-5).
Pressing the or button (Fig. 3-7 and 9)
increases or reduces the speed value by 1 rpm.
Pressing and holding the or button (Fig. 3-7
and 9) increases or reduces the speed value.
If the torque or speed value is selected wrongly,
!min! or !max! will be displayed, followed up by
the maximal or minimal value permitted.
e) Torque adjustment
Press button . The cursor is placed under the
first figure.
Adjust the value of the first figure using the or
button (Fig. 3-7 and 9).
Press button . The cursor is placed under the
next figure.
Repeat this sequence figure by figure.
Confirm the torque value by pressing button .
f) Footswitch type selection
When the torque has been adjusted, the cursor is
displayed under the footswitch symbol.
Press button to toggle between ON/OFF and
progressive.
g) Completion of setting adjustment
To be able to use your device with the new
settings, press button .
The storing of the various settings is confirmed by
an audible signal and by the symbol which
blinks on the screen.
Repeat this sequence if necessary in order to
adjust all the programs.
7. 5 PIEZOTOME FUNCTION
Important:
The P
IEZOTOME
function power settings cannot
be adjusted by the user. Only the irrigation
settings can be user-defined.
By default, the device is in P
IEZOTOME
mode
when no ultrasonic handpiece is connected to
it.
The spray for the P
IEZOTOME
function can only be
adjusted when a P
IEZOTOME
handpiece is
connected to the device.
Select the P
IEZOTOME
function by pressing the right
Selection button (Fig. 3-10).
On the LCD screen, the P
IEZOTOME
tab must be
highlighted (Fig. 6).
Select the program to be used , , or (Fig.
3-5); the selected program is displayed on the
LCD screen.
To confirm the choice of program, press button
.
If necessary, modify the Irrigation function (flow
rate value, etc.)(refer to 6.3).
Confirm the new settings by pressing button .
13