44
Specifications
Feature Specification
Measurement Range
20 - 600 mg/dL
Result Calibration
Plasma-equivalent. Calibrated by using YSI (Model 2300
STAT PLUS) Glucose Analyzer reference instrument, which
is traceable to NIST reference standard.
Sample
Fresh capillary whole blood
Minimum Sample Size
0.8 µL
Test Time
5 seconds
On/Off Source
Two (2) CR 2032 3.0V coin cell batteries
Battery Life
1,000 tests for glucose testing (not considering data
transfer). Meter display backlight and strip port light are
powered by a separate battery.
Glucose Units of Measure
The meter is preset to either millimoles per liter (mmol/L) or
milligrams per deciliter (mg/dL) depending on the standard
of your country.
The meter will be set to mg/dL by default
when sold in the United States.
Memory
Up to 500 records with time and date
Automatic Shutoff
2 minutes after last action
Meter Size
89.6 mm × 58.0 mm × 21.7 mm
Display Size
37.0 mm × 33.5 mm
Weight
60.0 g (with batteries installed)
Operating Temperature
41 – 113°F (5 – 45ºC)
Operating Relative Humidity
10 – 90% (non-condensing)
Hematocrit Range
20 – 70%
Data Port
9600 baud, 8 data bits, 1 stop bit, no parity