35
Maintenance
Proper maintenance is recommended for best results.
Replacing the Battery
Your
On Call
®
Vivid
Meter uses two 3.0 Volt CR 2032 lithium batteries. One battery
(Meter Battery) powers only the meter. The other battery (Backlight & Strip Port
Light Battery) powers only the meter display backlight and test strip port light.
When the battery icon ( ) is
blinking, it means the Meter Battery
is running low. You should replace
the Meter Battery as soon as
possible. An “E-6” error message will
appear if the Meter Battery is too low
to perform any more blood glucose
tests. The meter will not function
until the Meter Battery is replaced.
Instructions:
1. Turn the meter off before removing the battery.
2. Turn the meter over to locate the battery cover. Press the battery cover tab on
the top. Lift the cover to get to the battery.
3. Remove and discard the old Meter Battery. Insert one new CR 2032 3.0V coin
cell battery on top of plastic tape. Make sure the plus (+) side facing up.
4. Close the battery cover and make sure that it snaps shut.
5. Recheck and reset the clock setting after battery replacement, if necessary.
To set the meter clock, see
Meter Setup Before Testing
on page
12
.
Low backlight and strip port light battery
Replace the Backlight and Strip Port Light Battery when you notice the light
dimming. The meter does not display a warning when there is a low backlight.
Note:
Having a low backlight will not affect the accuracy of the meter’s results.
Meter Battery
Backlight & Strip
Port Light Battery