20
the foil pouch).
•
Check the expiration date of the test strip and control solution. Make sure
that the test strips vial and control solution bottle have not been opened for
more than 6 months. Discard any test strips or control solution that has
expired.
•
Confirm the temperature in which you are testing is between 50 and 104°F
(10-40°C).
•
Make sure that the test strips vial and control solution bottle have been tightly
capped.
•
Confirm that you are using the same brand of control solution that was
provided with your kit.
•
Make sure that you followed the test procedure correctly.
After checking all of the conditions listed above, repeat the quality control test with
a new test strip. If your results still fall outside of the control range shown on the
test strip vial (or on the foil pouch), your meter may be defective. Please contact
Customer Support at 1-800-838-9502.
Two levels of control solution are available labeled Control Solution 1 and Control
Solution 2. Control Solution 1 is sufficient for most all self-testing needs. If you think
your meter or strips may not be working correctly, you may also want to do a level 2
test. The ranges for both (CTRL 1 and CTRL 2) are displayed on the test strip vial
(or on the foil pouch). Simply repeat step 4 through 6, using Control Solution 2.
To confirm your results, Control Solution 1 tests should fall within the CTRL 1 range,
and Control Solution 2 tests should fall within the CTRL 2 range. If the control
solution test results do not fall within their respective ranges, DO NOT use the
system to test blood, as the system may not be working properly. If you cannot fix
the problem, please contact Customer Support at 1-800-838-9502.
Please contact Customer Support at 1-800-838-9502 for information on ordering
the
On Call
®
Vivid
Glucose Control Solution kit. The kit contains Control Solution 1
and Control Solution 2.
Содержание On Call Vivid