To remove dust, dirt and other impurities not originating from the patient, wipe clean with a
soft, lint-free cloth moistened with ethanol or 70 % isopropyl alcohol. When cleaning, wear
gloves for protection from chemical products.
Ensure that the equipment is completely dry before use.
Use soft non-woven cloths to clean the lenses so as to avoid scratching them.
9 After-sales service and maintenance
This device does not require any special maintenance operations.
Note:
Misuse of the device is not covered by the warranty.
If a fault persists and the device has to be returned to the distributor, ensure that it is shipped
in its original packaging.
Similarly, it is advisable to return the device in its entirety (control unit and power cables).
Kindly attach to the shipping order a short explanatory note about the fault detected.
The equipment must be disinfected before it is returned for repair. When returning the
equipment, check its condition and note any exceptions on the shipping order if necessary,
and confirm these with the shipper by registered mail as soon as possible.
Please contact your distributor who will guide you through the return process.
10 Troubleshooting
10.1 The indicator light on the standby switch S1 does not light up on
start-up
•
Check that the mains socket at the back of the camera is properly connected to the power
supply and that the general switch at the back of the camera is in the operating position
(“I”).
•
Check that the fuses are in good condition (only use the fuses specified on the back panel).