USER MANUAL 16-2040N / 16-2041N / 16-2042N
Rev. 07/2016
11. ELECTROMAGNETIC COMPATIBILITY
Guide and declaration by manufacturer – electromagnetic emissions
This insufflator is designed to be used within the electromagnetic environment specified below. The user should ensure
that it is indeed used within this environment
EMISSIONS TEST
COMPLIANCE ELECTROMAGNETIC ENVIRONMENT-GUIDE
RF emissions
CISPR 11
Group 1
This insufflator only uses radioelectrical energy for its internal
subsystems. Therefore, it emits very low RF energy and is not likely to
interfere with nearby electronic devices.
RF emissions
CISPR 11
Class A
This insufflator must be used in all premises other than residential
premises and premises connected directly to the public low-voltage power
distribution network used for powering residential buildings.
Harmonic emissions
EN 61000-3-2
Conform
Voltage fluctuations /
flicker EN 61000-3-3
Conform
Guide and declaration by manufacturer – electromagnetic immunity
This insufflator is intended to be used within the electromagnetic environment specified below. The user should ensure
that it is indeed used within this environment
IMMUNITY
TEST
CEI 60601
SEVERITY LEVEL
COMPLIANCE
LEVEL
ELECTROMAGNETIC ENVIRONMENT-GUIDE
Electrostatic
discharges
EN 61000-4-2
± 6 kV via contact
± 8 kV via air
± 6 kV
± 8 kV
The floor should be in timber, concrete or tiling. If the
floor is covered with a synthetic material, the relative
humidity should be at least 30%.
Quick transients
in bursts
EN 61000-4-4
± 2 kV for power lines
± 1 kV for
input/output lines
± 2 kV
± 1 kV
The quality of the main power supply should be that of
a typical commercial or hospital environment.
Voltage shocks
EN 61000-4-5
Differential mode
± 1 kV
Common mode ± 2 kV
± 1 kV
N.A.
The quality of the main power supply should be that of
a typical commercial or hospital environment.
Brown-outs, brief
power
failures and
voltage
variations
EN 61000-4-11
• <5% U
T
– for 10 ms
• 40% U
T
– for 100 ms
• 70% U
T
– for 500 ms
• <5% U
T
– for 5 s
<5% U
T
10 ms
<40% U
T
100 ms
<70% U
T
500 ms
<5% U
T
5 s
The quality of the main power supply should be
that of a typical commercial or
hospital environment.
If the user of this insufflator requires being able to
continue to work during interruptions in the main
power supply, it is recommended that this insufflator
be powered via a UPS or battery.
Magnetic field at
network
frequency
(50/60 Hz)
3 A/m
3 A/m
The magnetic field at the network frequency should be
at a level characteristic of a location within a typical
commercial or hospital environment.
Note: U
T
is the nominal value of the power voltage applied during the test.
PAGE 19