![ACKERMANN 16-2040N Скачать руководство пользователя страница 16](http://html1.mh-extra.com/html/ackermann/16-2040n/16-2040n_user-manual_458090016.webp)
USER MANUAL 16-2040N / 16-2041N / 16-2042N
Rev. 07/2016
7. FAULTS AND WARNING SIGNALS
WARNING MESSAGE
POSSIBLE CAUSE
ACTION TO BE TAKEN
CYLINDER LOW
Cylinder pressure at 40 bar
Prepare a replacement cylinder
Cylinder pressure at 20 bar
Change the cylinder immediately
Gas cylinder valve closed
Open the valve
Insufflator high-pressure hose connector
disconnected
Turn off the device, check CO
2
intake
and restart the device
OVER-PRESSURIZATION
Sporadic action on pneumoperitoneum
None (if pressure is 5 mmHg greater
than blow-off pressure for more that 3
seconds the device performs an
external desufflation)
Tubing kinked
Stretch the tubing
TUBING NOT THERE
The tubing is missing
Connect the tubing
The tubing is wrongly connected
(desufflation actuator)
Connect the tubing correctly
CALIBRATION ERROR
The tubing is connected
Disconnect the tubing from
insufflator’s outlet connector for 2
seconds only
REMOVE TUBING
AUTOMATIC TEST not run
Remove the tubing for 2 seconds,
then replace
Fault in tubing detection
Send device for servicing
CO
2
NOT DETECTED
Gas cylinder valve closed
Open the valve
Insufflator high-pressure hose
disconnected
Connect the hose correctly to the
insufflator
SERVICE
Maintenance action essential
Send device for servicing
No power
Fuses damaged
Check and replace the 2.5A UR fuses
if necessary
For any other problem, please contact your nearest after-sales service team or our HQ in Germany.
8. CLEANING / SERVICE / MAINTANCE
8.1 CLEANING / DECONTAMINATION
Only instructions created by the personnel responsible for sterilization in each hospital or health
centre must be followed; these instructions will take precedence over the information contained in
this manual, which is provided for guidance only.
Always disconnect the device’s electrical plug before cleaning it.
After each use:
• discard disposable sterile tubing; do not try to sterilize it
• clean up any possible splashes of liquid on the insufflator by wiping them with a slightly dampened cloth.
The device MUST be decontaminated before it is sent to the after-sales service team.
PAGE 16
OF 21