The AUTOMATIC TEST will only run if tubing is not connected. It only takes a second.
You may now connect the tubing as described
in the following chapter
The device is waiting for the blow-off
pressure and will not start until it receives
this information.
6.2 CONNECTING-DISCONNECTING THE PATIENT TUBING
We cannot be held liable for any failure or malfunctions arising from the use of damaged or
unsuitable tubing.
Do not use tubing if the packaging is damaged.
Sterile tubes are disposable; do not re-sterilize.
Use of a hydrophobic bacterial filter is essential for the prevention of cross-contamination with the
patient.
Open the first set of packaging film around the tubing. Hand the second set to sterile personnel.
Keep the patient’s sterile tube in the operating theatre.
Open the second sterile pouch and connect the tubing to the insufflator’s gas outlet.
Insert the other part of tubing into the external desufflation actuator
The tube needs to be straight, inserted with ease and must not be obstructed. Do not switch the two
tubes, as they have different diameters.
• Connect the Luer lock to the trocar port on the patient side.
PAGE 11