![ACKERMANN 16-2040N Скачать руководство пользователя страница 15](http://html1.mh-extra.com/html/ackermann/16-2040n/16-2040n_user-manual_458090015.webp)
USER MANUAL 16-2040N / 16-2041N / 16-2042N
Rev. 07/2016
6.10 MONITORING THE VOLUME OF GAS AVAILABLE
As soon as the cylinder pressure reaches 40 bars a flickering cylinder icon is displayed.
At 20 bars a red light alarm light [B] and an alarm message “CYLINDER PRESSURE 20 bar” is
displayed.
The cylinder status in the display is updated continuously.
Below 10 bars it is not possible to start the device (RUN/STOP key disabled); the cylinder must be
changed immediately.
No operation may proceed using a cylinder below a level which would not last the duration of the
surgery.
6
.
11
END OF OPERATION
Stop insufflation by pressing
[F]
; the STOP icon appears.
Disconnect the tubing from the trocar and the insufflator immediately to prevent any liquid or gas from flowing
back to the device.
The used tubing must be disposed of in an appropriate container after use.
The total volume of gas insufflated during the operation appears; reset the gas meter to zero if another
operation is planned (see chapter VII – C).
Close the CO2 gas cylinder
PAGE 15
OF 21