Detection of Broken Navvus® MicroCatheter
40
| ACIST RXi System User’s Guide
901700-001,01 2019-09 English
Basic Operating Procedures
WARNING
If a broken fiber optic is detected within the Navvus MicroCatheter, a visual and auditory
technical alarm condition is triggered. The defective microcatheter, along with the guide
catheter, must be immediately removed and replaced.
The Catheter Malfunction alarm, distinguishable from all other system
information signals, sounds immediately, and consists of a series of repeated
tones at a maximum volume of 80 dB ±5%. (Other auditory signals are a single
short pulse at low volume.) Simultaneously, the following flashing message is
displayed:
Warning!
Catheter malfunction detected.
Stop procedure immediately and replace catheter with a new one.
The alarm continues to sound and the message displayed until the defective
Navvus MicroCatheter is disconnected.
Detection of
Broken Navvus®
MicroCatheter