EMC Tables
104
| ACIST RXi System User’s Guide
901700-001,01 2019-09 English
Technical Specifications
Essential Performance
Measure blood pressure with
Navvus® MicroCatheter. The pressure
static accuracy shall be ±3% of
reading or ±3 mmHg of reading over
operating range of -30 mmHg to 300
mmHg.
The RXi System to begin alarming
when a broken fiber optic condition
occurs.
Immunity Tests
During test, the system is set to read a positive value (~20 mmHg on the
screen) with a 0.5 Hz square 5 mmHg wave imposed on it. The pressure
is monitored during testing to ensure compliance with the Essential
Performance.
Electrostatic Discharge (ESD)
4th Edition: Criterion A
The system continuously monitors for
a broken fiber. No specific testing was
performed during EMC-EMI testing to
validate broken fiber detected during
testing.
Electrical Fast Transient / Burst 4th Edition: Criterion B
Surge
4th Edition: Criterion A
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
4th Edition: Criterion A
4th Edition: 250/300 cycle dip
Criterion C
4th Edition additional Phase angles
Criterion A
Power Frequency (50/60 Hz)
Magnetic Field
4th Edition: Criterion A
Conducted RF Susceptibility
Criterion A
Radiated RF Susceptibility
Criterion A
Radiated RF
4th Edition spot
Criterion A
Emissions Test
RF Emissions
Passes class B emissions* per TÜV SÜD America, Inc.
NOTE: The RXi System uses RF energy only for its internal function. Therefore,
its RF emissions are very low and are not likely to cause any interference in
nearby electronic equipment.
RF Emissions Conducted
Class B
* The ACIST RXi® System is in compliance with Class A RF emissions when the Ao Interface Box is
attached.