Basic Operating Procedures
ACIST RXi System User’s Guide |
33
901700-001,01 2019-09 English
Unpack and Prepare the Navvus® MicroCatheter
Preparation Needed
The following instructions provide technical direction, but do not eliminate
the need for formal training in the use of the Navvus MicroCatheter and
the ACIST RXi System. The techniques and instructions provided here do
not represent ALL medically acceptable protocols, nor are they intended as a
substitute for the physician’s experience and judgment in treating any specific
patient.
Required Materials
Each Navvus MicroCatheter package contains the following components:
• one single-use disposable microcatheter
• one dispenser coil
• one monorail stylet
Other materials that are required
but not supplied
include the following:
• 5F to 8F guide catheter
• standard guidewire
• hand syringe used for flushing the dispenser coil and catheter
• sterile saline solution
CAUTION
To avoid inaccurate aortic pressure measurements, the use of guide catheters larger
than 8F or guide catheters with side holes are not recommended. Avoid blocking blood
flow to the selected artery. Do not wedge the guide catheter in the selected artery.
WARNING
Prior to use and whenever possible during the procedure, carefully inspect the Navvus
MicroCatheter for kinks or any other damage. Do not use a kinked or damaged
microcatheter because vessel damage and/or inability to advance or withdraw the
microcatheter may occur.
The Navvus MicroCatheter is supplied sterile. If the sterile package is damaged or the
seal breached, discard the product. Do not re-sterilize the microcatheter. Do not use the
product beyond the expiration date indicated on the product packaging.
Improper technique may damage the Navvus MicroCatheter. Exercise care in handling
the microcatheter during a procedure to reduce the possibility of accidental breakage or
kinking.
Unpack and
Prepare the
Navvus®
MicroCatheter
Содержание RXI
Страница 1: ...User s Guide ACIST RXi Rapid Exchange FFR System Bracco Group 901700 001 01 2019 09 English ...
Страница 10: ...4 ACIST RXi System User s Guide 901700 001 01 2019 09 English Introduction This page intentionally left blank ...
Страница 123: ......
Страница 124: ...901700 001 01 2019 09 English ACIST Medical Systems Inc 7905 Fuller Road Eden Prairie MN 55344 USA ...