Installation
ACIST RXi System User’s Guide |
25
901700-001,01 2019-09 English
Installation Procedures
Zero the Aortic Pressure on the ACIST RXi® System
To ensure that the hemodynamic monitoring system is reading a zero value:
• Turn on power to the system. (Refer to page 30 for instructions.)
• With the stopcock at the midaxillary position, open the stopcock to
expose the hemodynamic sensor to atmospheric pressure at the patient
midaxillary position.
• Zero the hemodynamic monitor.
Zero the aortic pressure on the ACIST RXi System:
1. When the following status message appears on the touchscreen, press Zero
Aortic:
Zero Aortic Pressure
Press
Zero Aortic
at the system
prompt.
2. Verify that the aortic signal (Pa) is showing zero (0) in the display area.
Also verify that the hemodynamic monitor is showing zero (0).
In this
example, the
aortic signal
(Pa)
is zero.